Series 500 Portable Gas Recorder |
|
|
|
_WRITTEN_BY Administrator
|
|
_DATE_FORMAT_LC |
|
Data Recorder function ,max 8,000 datas, over 20 varieties Gas Sensor for Aeroqual Series 500
| Series 500 Monitor Base |
| Measurement units |
ppm or mg/m 3 |
| Removable / replaceable sensor head |
Yes |
| On-board alarm |
Yes |
| Alarm status displayed |
Yes |
| Low battery indication |
Yes |
| Alarm mute function |
Yes |
| Sensor condition status |
Yes |
| Stand-by mode |
Yes |
| External signal for alarms & control |
Transistor output, 150 mA max |
| External signal functions |
Low Alarm, High Alarm & Control |
| Analogue output |
0 - 5 V |
| Power supply |
12 VDC, 800 mA |
| Rechargeable battery pack (optional) |
9.6V Ni-MH (5 hours operation) |
| Portable (with optional battery pack) |
Yes |
| Permanently fixable |
Screw fix |
| Remote Sensor Capability |
Yes |
| Enclosure rating |
IP20 & NEMA 1 equivalent |
| Size (with sensor head) |
195 x 122 x 54 (mm); 7 x 4 x 2 (in) |
| Weight (with sensor head and battery) |
< 460 g; < 16 oz |
| Approvals |
Part 15 of FCC Rules
EN 50082-1: 1997
EN50081-1: 1992 |
| Data interface with PC |
Serial RS232 |
| On Board Data Logging |
8,000 data points |
| Data logging independent of PC |
Yes |
| Data logging direct to PC |
Yes |
| On-board real-time clock |
Yes |
| Data logging software supplied |
Yes |
Ozone Sensor Head Specifications
| Sensor Heads |
Low |
High |
Leak |
| Measurement range |
0.000 - 0.500 ppm |
0.50 - 20.00 ppm |
0.00 - 50.00 ppm |
| Accuracy |
< 0.008 ppm (0 - 0.100 ppm)
± 10% (0.100 – 0.500 ppm) |
± 10% (0.20 - 2.00 ppm)
± 15% (2.00 - 20.00 ppm) |
± 20%, 15 seconds after Reset |
| Resolution |
0.001 ppm |
0.01 ppm |
0.01 ppm |
| Controlled Atmosphere Version |
Available on Request |
Available on Request |
Available on Request |
| T90 Response |
< 60 seconds (T90) |
< 35 seconds (T90) |
< 10 seconds |
| Sensor type |
Gas-sensitive semiconductor |
| Operating temperature range |
-5°C to 50°C; 23°F to 122°F (sensor head and base unit) |
| Relative humidity limit |
95% maximum (sensor head and base unit) |
Ammonia Sensor Head Specification
| Measurement range |
0 to 1000 ppm |
| Accuracy |
< +/- 5 ppm (0 – 100 ppm)
< +/- 10% (100 – 1000 ppm) |
| T90 response |
< 60 s |
| Sensor type |
Gas-sensitive semiconductor |
| Operating temperature range |
-20°C to 60°C; |
| Operating relative humidity range |
5 to 95% non-condensating |
CO Sensor Head Specification
| Measurement range |
0 to 1000 ppm |
| Accuracy |
< +/- 5 ppm (0 – 100 ppm)
< +/- 10% (100 – 1000 ppm) |
| T90 response |
< 60 s |
| Sensor type |
Gas-sensitive semiconductor |
| Operating temperature range |
-20°C to 60°C; |
| Operating relative humidity range |
5 to 95% non-condensating |
|
|
_LAST_UPDATED ( _DATE_FORMAT_LC )
|